Technical Parameters Of Micro Ohm Meter :-
Power Supply : 230VAC,± 15%, 50Hz(± 5%)
Source : 100 A DC Regulated source (In settings 10,20,50,100A)
Measuring Current : 100ADC
Measuring Range :1mW to 1999 mW
Resolution : 0.1µ Ohm (at lowest range)
Accuracy : ± 2% of reading
Display : 2 Nos. 3 ½ DIGIT, LCD DISPLAY or LED Display.
LED indication for over Temp. Open circuit Voltage : 30V DC.upto 20Amp,12VDC above 20Amp.
Operating Temp :10 to + 50 Degree Centigrade.
|